János, Tóth F. wrote: > What is the valid range for the target integration time parameter? > > 0.1-4.0 comes to my mind from somewhere (may be CalMan?) but I could > not find a word about this in the ArgyllCMS documentation. Hi, It's whatever I want it to be, within the hardware counter limits (ie. there will be a maximum edge count in frequency measurement mode that therefore depends on the integration time and the light level, and there will be similar limits for the period measurement mode.) By default the integration time aimed at is 0.2 seconds. It will take longer than that for low light levels if that's needed to measure one period of the L2F. I was thinking about making -yr (r for refresh) set it to 0.4 seconds, halving the maximum measurement speed. Graeme Gill.