Thanks Paul.  All good tips

Denice

On Tue, Apr 9, 2024 at 12:10 PM Paul Crown (fotoneurotic) <
fotoneurotic@xxxxxxxxx> wrote:

It's great for video angle, terrible for eyelashes.

If you angle the patients chair instead of changing the angle of their
neck....they tend to fight positioning less.

Everyone tries to put their nose in it.

Patients pick up the habit of the technician at their local optometry
office

Green doesn't always mean it's in the right distance position.  Sometimes
the red side of green or the blue side of green is better than green.

It's bright, but not as bright as the Clarus

Don't make patient info changes on the tablet at the scan head, only ever
make changes within OptosAdvance itself.  Changes in the tablet create
duplicates seen on the tablet and doesn't connect the previous info to the
same patient every time or can connect the patient to a different patient.

It's slow on FA's.  A pic every 4 seconds on FA, FA/ICG combined is 1 Pic
of a modality every 8 seconds so 4 sec FA 4 sec ICG. 4 sec FA 4 sec ICG.
Unless you shoot separate transits, but then it's still an image every 4
seconds.  Transits are the issue there.

Everyone will grab the table.

Wheelchairs pull up nice, may not always need to move the foot rests out.


Sent via the Samsung Galaxy S22 Ultra 5G, an AT&T 5G smartphone
Get Outlook for Android <https://aka.ms/AAb9ysg>
------------------------------
*From:* optimal-bounce@xxxxxxxxxxxxx <optimal-bounce@xxxxxxxxxxxxx> on
behalf of Denice Barsness <denicebars@xxxxxxxxx>
*Sent:* Tuesday, April 9, 2024 9:40:01 AM
*To:* optimal@xxxxxxxxxxxxx <optimal@xxxxxxxxxxxxx>
*Subject:* [optimal] Optos

Any superuser tips on the pros/cons of using this instrument?  Any tips
and tricks?

Many thanks
Denice Barsness