Am 22.11.2019 um 15:26 schrieb Vallo Kallaste:
Hmm, this is contradictory. The same instrument+device combo in one case
gives data so faulty that you warn user and err out. Very good approach.
Yet in another case the same instrument+device combo provides data good
enough to proceed.
Nothing contradictory here. This can happen if (including but not
limited to):
- Display has any "dynamic" functionality enabled (dynamic contrast,
auto brightness, etc.)
- "Dynamic" system features like Night Shift, True Tone or other
software interfering
- Stray ambient light entering the sensor due to lighting not being
consistent (e.g. you switch on/off a bright room light during measurements)
- Monitor defect (unlikely)
- Instrument defect (unlikely)
--
Florian Höch