[argyllcms] Re: Wanted: Spyder 3 and ColorMunki Beta Testers - source update

  • From: Nikolay Pokhilchenko <nikolay_po@xxxxxxx>
  • To: argyllcms@xxxxxxxxxxxxx
  • Date: Sun, 28 Jun 2009 14:17:32 +0400

I've noticed the difference between chartread 1.0.3 and 1.1.0_b versions in 
high resolution spectrum mode for i1 Pro instrument. The older version gives 
113 spectral bands, the newer - only 112:

1.1.0_b version:
KEYWORD "TARGET_INSTRUMENT"
TARGET_INSTRUMENT "GretagMacbeth i1 Pro"
KEYWORD "SPECTRAL_BANDS"
SPECTRAL_BANDS "112"
KEYWORD "SPECTRAL_START_NM"
SPECTRAL_START_NM "376.890756"
KEYWORD "SPECTRAL_END_NM"
SPECTRAL_END_NM "750.000000"

1.0.3 version:
KEYWORD "TARGET_INSTRUMENT"
TARGET_INSTRUMENT "GretagMacbeth i1 Pro"
KEYWORD "SPECTRAL_BANDS"
SPECTRAL_BANDS "113"
KEYWORD "SPECTRAL_START_NM"
SPECTRAL_START_NM "376.666667"
KEYWORD "SPECTRAL_END_NM"
SPECTRAL_END_NM "750.000000"

This difference prevents the combining spectral data from different versions 
*.ti3 files.

Florian HЖch wrote:

> Hello,
> 
> I compiled win32 binaries using MinGW/ftjam a while ago. You may get 
> them here:
> 
> http://hoech.net/files/Argyll_V1.10_Beta_win32.zip
> 
> Regards,
> 
> Florian HЖch
> 


Other related posts: