Niccolò Belli wrote:
Failed to meet target 0.400000 delta E, got worst case 0.362716
I've explained this once before: Each iteration tries to meet the target threshold. If it fails, it reverts to the previous best. The previous best was better than what was achieved at this iteration. (ie. the display and/or instrument aren't perfectly repeatable). Graeme Gill.