Hi Peter,
 
We had the very same problem and fixed it with a new firewire cable.
 
Sean Grout
 
Photo Dept Manager
West Coast Retina
185 Berry Street, ste 130
San Francisco, CA 94107
(415) 972-4634

--- On Tue, 5/21/13, John Horna <johnhorna@xxxxxxxxx> wrote:


From: John Horna <johnhorna@xxxxxxxxx>
Subject: [optimal] Re: OIS WS-5000 problem
To: optimal@xxxxxxxxxxxxx
Date: Tuesday, May 21, 2013, 10:33 AM



My educated guess would be static electricity.
 
Look in Device Manager (assume Win XP), and the camera is likely unrecognized, 
but a reboot will make it OK again.  
 
Perhaps if you can ground it, it may eliminate the problem.  You can also try 
diconnecting and reconnecting Firewire to the computer prior to reboot, but I 
would not hold my breath on that one.
 
John


On Tue, May 21, 2013 at 9:24 AM, Breit, Peter <BreitP@xxxxxxxx> wrote:




Dear colleagues 
 
The Fundus camera is a Topcon TRC-50 EX serial # 1350878
Digital camera is a Megavision 5M model FB-3020 10M (got this information from 
the list of devices attached to the computer) – the number on the camera back 
is 28R015455000
OIS is version 10.5.74.0 – the disc is dated 5/7/08 but I am not sure of the 
date that we installed it.
 
I am having camera (WS-5000) problems, I believe it is the connection of the 
Firewire to the camera. The Firewire is new but as soon as it touch it 
inadvertently the computer cannot find the camera and gives me the following 
error codes:
 

“ Too many errors reentering capture cycle attempting to start camera WS-5000 
Color”
“ Attempt to reinitialize camera WS-5000 Color failed. System now will stop 
capture using this camera until    
  program restarted and problem resolved. “
 
I don’t believe it is a software problem, but I am not sure. There is a lot of 
other stuff loaded on the computer for             e-records, etc.
 
Any help is very much appreciated. Before I send the camera to Megavision for 
repair. 
 
 
Thank you
Peter
 
Peter L. Breit, CRA.
Director Ophthalmic Services
The Lankenau Hospital & Bryn Mawr Hospital
484-476-3338
484-476-8206 fax
484-437-3262 mobile 
page 3707
 
 
 
 
 
 

Other related posts: