[qudi-announce] Changes in Ni-card config

  • From: nikolas.tomek@xxxxxxxxxx
  • To: qudi-announce@xxxxxxxxxxxxx
  • Date: Thu, 01 Feb 2018 11:39:19 +0100

Hello everyone,

since today it is possible to use analog input channels from e.g. a photo diode instead of an APD for counting.

The current ni-card module was renamed to "national_instruments_x_series.py" so people do not wrongly assume that M-Series cards should work fine and then complain.

This leads to some changes in the config for NI card.

Old configuration:

mynicard:
     module.Class: 'ni_card.NICard'
     clock_channel: '/Dev1/Ctr0'
     counter_channel: '/Dev1/Ctr1'
     scanner_clock_channel: '/Dev1/Ctr2'
     scanner_counter_channel: '/Dev1/Ctr3'
     photon_source: '/Dev1/PFI8'
     photon_source2: '/Dev1/PFI9'
     gate_in_channel: '/Dev1/PFI9'
     counting_edge_rising: True
     odmr_trigger_channel: '/Dev1/PFI15'
     scanner_x_ao: '/Dev1/AO0'
     scanner_y_ao: '/Dev1/AO1'
     scanner_z_ao: '/Dev1/AO2'
     scanner_a_ao: '/Dev1/AO3'
     clock_frequency: 100
     scanner_clock_frequency: 100
     x_range:
     - 0e-6
     - 200e-6
     y_range:
     - 0e-6
     - 200e-6
     z_range:
     - -100e-6
     - 100e-6
     a_range:
     - -10
     - 10
     voltage_range:
     - -10
     - 10
     samples_number: 10

New configuration:

   mynicard:
module.Class: 'national_instruments_x_series.NationalInstrumentsXSeries'
        clock_channel: '/Dev1/Ctr0'
        scanner_clock_channel: '/Dev1/Ctr2'
        photon_sources:
            - '/Dev1/PFI8'
            - '/Dev1/PFI9'
        counter_channels:
            - '/Dev1/Ctr1'
        counter_ai_channels:  # optional
            - '/Dev1/AI1'
        scanner_counter_channels:
            - '/Dev1/Ctr3'
        scanner_ai_channels:  # optional
            - '/Dev1/AI0'
        scanner_ao_channels:
            - '/Dev1/AO0'
            - '/Dev1/AO1'
            - '/Dev1/AO2'
            - '/Dev1/AO3'
        scanner_position_ranges:
            - [0e-6, 200e-6]
            - [0e-6, 200e-6]
            - [-100e-6, 100e-6]
            - [-10, 10]
        scanner_voltage_ranges:
            - [-10, 10]
            - [-10, 10]
            - [-10, 10]
            - [-10, 10]
        default_samples_number: 10
        default_clock_frequency: 100
        default_scanner_clock_frequency: 100
        gate_in_channel: '/Dev1/PFI9'
        counting_edge_rising: True
        odmr_trigger_channel: '/Dev1/PFI15'


Cheers
Nikolas

--
Nikolas Tomek
Institute for Quantum Optics
Ulm University
Albert-Einstein-Allee 11
89081 Ulm
Germany

Phone: +49 731 50-15711
E-Mail: nikolas.tomek@xxxxxxxxxx






Other related posts:

  • » [qudi-announce] Changes in Ni-card config - nikolas . tomek