You need to set up a network share for the Data folder on the Data drive (Q, in our case), and then map that share to a drive letter on each review station. Then, when installing the review software, you will be asked for the path to the data. Simply set it up to look at the mapped drive for the data. We not only have multiple review stations looking at the same database, but, because we have multiple Spectralis units in multiple offices, I have even set up a few computers to have multiple instances of the review software that point to all the different machines. You are limited to the number of active review stations specified by your license. George E. Henry, CRA, PBT Wheaton Eye Clinic, Ltd. 2015 N. Main Street Wheaton IL 60187 630.588.3615 _____ From: optimal-bounce@xxxxxxxxxxxxx [mailto:optimal-bounce@xxxxxxxxxxxxx] On Behalf Of Tom Steele Sent: Wednesday, July 17, 2013 06:03 To: optimal@xxxxxxxxxxxxx Subject: [optimal] Heidelberg Review Stations Help Everyone, How do you handle multiply review stations for the Heidelberg within the clinic? Right now the doctors want to be able to open more than 1 review station at a time from multiply exam rooms; thus I might be imaging a patient and 3 doctors want to review 3 different patients at the same time. Perhaps someone from Heidelberg will be reading this and can point a solution to me. Thank You, Tom Steele, CRA Midwest Eye Institute 200 West 103rd Street Indianapolis, Indiana 46290 317.817.1018 tsteele@xxxxxxxxxxxxxx CONFIDENTIALITY NOTICE: This message is intended only for the use of the individual or entity to which it is addressed and may contain information that is confidential and/or privileged. If the reader of this message is not the intended recipient, you are hereby notified that any dissemination, distribution or copying of this communication is strictly prohibited. If you have received this communication in error, please notify us immediately by telephone and return the original message to us at the above email address _____ No virus found in this message. Checked by AVG - www.avg.com Version: 2013.0.3349 / Virus Database: 3204/6498 - Release Date: 07/17/13