So, I suppose you'll have a fix for everyone available in some not too distant future update? Roger > -----Message d'origine----- > De : argyllcms-bounce@xxxxxxxxxxxxx [mailto:argyllcms- > bounce@xxxxxxxxxxxxx] De la part de Graeme Gill > Envoyé : 12 février 2009 02:34 > À : argyllcms@xxxxxxxxxxxxx > Objet : [argyllcms] Re: i1pro wrong XYZ values > > Jörg Koch wrote: > > > I measured color patches on my notebook LCD using the i1pro and > > experienced much higher absolute XYZ values when comparing these > > measurements with the measurments using the i1pro x-rite drivers + > > software. When measuring with the sypder3 using spyder3 software I > get > > similar values than the i1pro with the x-rite software, thus there > seems > > to e a problem with the Argyll measurements. > > Hmm. Well, it took a bit of effort, but I think I've finally > tracked this one down. There were three separate bugs, covering > each other up, involving the spectral to XYZ conversion. > Currently the emissive XYZ values are too big for the i1pro by > a factor of about 1.37 > > Graeme Gill.