Roger Breton wrote:
Is there an option, anywhere, that would allow the user to retain/save the absolute measurements, instead of normalized tristimulus values, where Y = 100? I can't find any.
You can do so with "dispread -w", if you want to get readings for a specific set of test values.
There must be an option to capture the spectral data too when it's an EyeOnePro?
"dispread -s" will add the spectral information. [There's no way of doing these things in dispcal, since its purpose is to output a calibration.] Graeme Gill.