Richard Hughes wrote:
Log file attached. Thanks.
munki: Read measurement results: inummeas 84, scanflag 0, address 0x998cf30 bsize 0x59e8 about to call usb_read with 23016 bytes Read 84 readings, ICOM err 0x0 (Trig & rd times 1 0 630 0) Needed scale is 0.837309 munki_optimise_sensor called, inttime 0.007168, gain mode 0, scale 0.837309 munki_optimise_sensor returning inttime 0.007168, gain mode 0 New inttime = 0.007168 Doing a white calibration with opt int_time 0.007168, gainmode 0
I'm not sure what's going on. 0.007168 is the minimum integration time, yet the result is too bright. For comparison, my instrument needs an integration time of 0.0182 (2.5 x more) to get close to the optimum white level. So either: There's something wrong with the sensor or LED. There's some bug in estimating the brightness level. So a lot more debug info is required to track down the problem. I'll take a look through the first trace (-D5) again and see if I can spot anything. Graeme Gill.