Thanks for the suggestion, but that wasn't it. I think I found the problem. In dispcalGUI, I'm waiting for the output "Hit ESC or Q to exit, any other key to take a reading" after sensor calibration, and this is where it times out, because the actual output with a ColorMunki has "instrument switch or" inserted before the "any other key" part. Well, atleast that should be easily fix-able by being less specific in the string I wait for :) Just "key to take a reading" should probably always work. I'll have a testing version ready later.
Regards Am 21.06.2010 17:08, schrieb Jürgen Lilien:
Jari Jylhä wrote:Yes it do. After pushing "Yes" button it says "Calibrating instrument..." and couts to 30 seconds. Then it says "Error".One dumb thought, maybe this problem is part of a misunderstanding. To start the calibration with argyll you have to press a button on the keyboard, not the button on the colormunki. Regards, Juergen
-- Florian Höch http://hoech.net