[argyllcms] Re: Can Eye-one Pro failure the calibration?

  • From: Graeme Gill <graeme@xxxxxxxxxxxxx>
  • To: argyllcms@xxxxxxxxxxxxx
  • Date: Tue, 05 Oct 2010 10:16:12 +1100

Sam Berry wrote:
Anyone have any ideas for the false patch 2 reading? This is repeatable.

Is this with dispread ?

One explanation would be that there is a synchronization problem.

The code sleeps for 60 msec after changing a color (which at
50Hz or more, is at least 3 refreshes), but perhaps the delay
with some cards, software and displays is longer than that ?

Another explanation would be that your display has some sort
of "auto contrast" adjustment mode on, and it takes some seconds
to react to image contents. Such a mode is (of course incompatible
with color management.

What operating system version are you running on ?

Graeme Gill.

Other related posts: