Roberto Michelena wrote:
4) make some sort of "Lab-to-Lab" link profile for each instrument to "match" the master one (DTP-70).
Hi, I'm not sure how well such a thing would work. It may work OK for a particular paper, ink & instrument pair, but it's hard to say. More serous corrections would work on spectral data, and the main complexity would be the allowance for illuminant differences interacting with FWA/OBA, where quite complex modeling would be needed (such as exists in the Argyll FWA compensation). In addition there may be geometric differences (where the nature of the light direction, detector direction sensitivity, and the relative sizes of the illuminated and detected regions differ), and I'm not sure how these sorts of differences manifest in the actual readings. Graeme Gill.