Richard Hughes wrote:
Using the 1.1.1 binaries from the argyll website:
Calibration failed with 'Measurement misread' (White reference calibration didn't converge) Hit any key to retry, or Esc or Q to abort:
I don't think anything in the ColorMunki driver has changed, so it's a bit of a puzzle. Perhaps the code is not forgiving enough in the real world. What does the debug output look like ? (-D5)
I can't seem to get the ColorMunki to self calibrate when using chartread, but it seems to work just fine when I'm using dispcal to calibrate the screen.
Emissive calibration skips the path you are having trouble with (setting an integration time to give optimal sensor levels). Graeme Gill.